+UHD - Calibration Application Notes
+========================================================================
+
+.. contents:: Table of Contents
+
+------------------------------------------------------------------------
+Self-calibration
+------------------------------------------------------------------------
+UHD comes with several self-calibration utilities for minimizing IQ imbalance and DC offset.
+These utilities perform calibration sweeps using transmit leakage into the receive path
+(special equipment is not required).
+The results from a calibration are written to a CSV file in the user's home directory.
+UHD will automatically apply corrections at runtime when the user re-tunes the daughterboard LO.
+Calibration results are specific to an individual RF board.
+
+**Note:**
+When a calibration table is present,
+and the user wishes to override the calibration settings through the API:
+the user should re-apply the desired setting every time the LO is re-tuned.
+
+UHD comes with the following calibration utilities:
+
+ * **uhd_cal_rx_iq_balance:** - mimimizes RX IQ imbalance vs. LO frequency
+ * **uhd_cal_tx_dc_offset:** - mimimizes TX DC offset vs. LO frequency
+ * **uhd_cal_tx_iq_balance:** - mimimizes TX IQ imbalance vs. LO frequency
+
+The following RF frontends are supported by the self-calibration utilities:
+
+ * WBX transceiver board
+ * SBX transceiver board
+ * RFX transceiver board
+
+********************************************
+Calibration Utilities
+********************************************
+UHD installs the calibration utilities into **<install-path>/bin**.
+**Disconnect** any extrernal hardware from the RF antenna ports,
+and run the following from the command line.
+Each utility will take several minutes to complete.
+::
+
+ uhd_cal_rx_iq_balance --verbose --args=<optional device args>
+ uhd_cal_tx_iq_balance --verbose --args=<optional device args>
+ uhd_cal_tx_dc_offset --verbose --args=<optional device args>
+
+See the output given by --help for more advanced options, such as:
+manually choosing the frequency range and step size for the sweeps.
+
+********************************************
+Calibration Data
+********************************************
+Calibration files are stored in the user's home/application directory.
+They can easily be moved from machine to another by copying the "cal" directory.
+Re-running a calibration utility will replace the existing calibration file.
+The old calibration file will be renamed so it may be recovered by the user.
+
+ * **Linux:** ${HOME}/.uhd/cal/
+ * **Windows:** %APPDATA%\\.uhd\\cal\\
+